AW-1 (Standard) infant radiant warmer
Brief introduction of infant radiant warmer:
--Microprocessor based servo controlled temperature system.
--The microprocessor-controlled unit allows for three settings: pre-warm, manual or automatic.
--Alarm functions of infant radiant warmer for safety includes:
Power failure;
Temperature deviation;
Temperature sensor failure;
over temperature.
--APGAR timer to record therapy time
--Convenient X-ray cassette under infant bed
--Heater head in infant radiant warmer can be adjusted +90º horizontally
--Tilt angle of infant bed is adjustable
--Observe light angle is adjustable
--Transparent protector folded
--Adopts anti-explosion micro crystal quartz infrared radiation tube for its radiation source, the heating is quick and life-span is long
--Alarms are audible and visual in infant radiant warmer
--A second thermal cut-out function for more safety
--Skin temperature sensor failure protect function avoid over temperature in infant radiant warmer
--Temperature deviation can be amended in front panel of infant radiant warmer directly
Standard configuration of infant radiant warmer:
Main body (including the radiant source, control system, infant bed, bracket), X-ray cassette , Observe light , I.V.pole, skin temperature sensor, tray, mattress, transparent protector, castors.
Optional parts of infant radiant warmer:
Drawer;
RS-232(485) connector;
Technical data of infant radiant warmer:
1. Power supply : 110/220V 60/50Hz
2. Power input: ≤1200VA
3. Operating condition:
An ambient Temperature:18°C-30°C
An ambient Relative Humidity: 30%--75%RH
Atmospheric Pressure: 700--1060 hPa
An ambient Air Velocity<0.3m/s
4. Temperature control: 32°C--38°C
5. Temperature control accuracy: ≤0.5°C
6. Mattress temperature uniformity :<2°C
7. Skin temperature sensor accuracy: + 0.3°C
8. Warm-up time: ≤30min
9. Infant bed tilt angle:+ 10°
10. APGAR timer range: ≤24h
11. Heater head horizontal: ±90°
12. Mattress size: 67cm × 54cm
13. Distance from heater to mattress:80cm
Transport and storage:
Environment Temperature:-40°C~ +55°C
Environment Relative Humidity: ≤95%
Atmospheric Pressure: 500--1060hPa
Package:
Each unit is packed one case
Size:120cm× 70cm × 96cm
Gross weight: ≤90kg
![]() |