Sign In | Join Free | My carsrow.com |
|
Place of Origin : China
Brand Name : ARI
Model Number : AW-2 (standard)
Payment Terms : T/T
Brief introduction:
--Three modes of control: manual,skin temperature and mattress temperature
--Alloyed bracket, fastness
--Microprocessor based servo controlled temperature system
--Control panel displayed by VFD
--Digital skin temperature sensor, the data is exact and stabilizability
--High sensitivity and untouched detector to inspect mattress temperature
--Alarms functions for safety includes
power failure temperature deviation
temperature probe failure over temperature
--Integral timer to record therapy time
--Observe light angle can be adjusted 360º, and the brightness adjusted
--Heater head could be adjusted+90º horizontally
--Electric infant bed tilt angle is+10º, transparent protector folded
--Convenient X-ray cassette under infant bed
--Electric height adjusted suit for operation
Standard configuration:
Main body (including the radiant source,control system,infant bed, electric lift stand),I.V.pole, skin sensor, mattress, one drawer, transparent protector, castors.
Function Features:
Adopts anti-explsion micro crystal quartz infrared radiation tube for its radiation source,the heating is quick and life-span is long
Alarms are audibility and visually
A second thermal cut-out function for more safety
Skin sensor failure protect function avoid over temperature
Technical data:
1.Power supply : 110/220V 50/60Hz
2.Power input: ≤1000VA
3.Operating condition
An ambient Temperature:18℃-30℃
An ambient Relative Humidity:30%--75%RH
Atmospheric Pressure: 700~1060Hpa
An ambient Air Velocity : <0.3m/s
4.Temperature control: 32℃~38℃
5.Temperature control accuracy: ≤0.5℃
6.Alarms functions:
a. Over temperatuer :>39℃
b. A sencond thermal cut- out : 40℃
c. Sensor alarm: skin temperature sensor failure
matterss temperature sensor failure
room temperature sensor failure
Power failure
7.Mattress temperature uniformity : <2℃
8.Skin temperature sensor accuracy: ≤0.3℃
9.Mattress temperature sensor accuracy: ≤0.4℃
10.Room temperature sensor accuracy: ≤0.3℃
11.Warm-up time: ≤40min
12.Height adjustable range:0-20CM
13.Bed tilt angle:+ 10°
14.Integral timer range: ≤24h
15.Heater head horizontal: ±90°
16.Mattress size:74CM X 64CM
17.Transport and storage
Environment Temp:-40℃~ +55℃
Environment Relative Humidity: ≤95%
Atmospheric Pressure: 500~1060Hpa
18.Whole unit size: 70CMX 110CMX 180CM(180-200CM)
19. Castor load: 1pc≤200kg 4pcs≤800kg
Package:
Each unit packed in one case:
Size:114*94*104cm GW<=190kg
![]() |
AW-2 (standard) Infant Radiant Warmer Images |